Vital Signs Simulator

Detailed Specifications

79

Respiration

Rate ........................................................................ 0 (OFF), 10 to 150 BrPM in 1 BrPM steps
Waves..................................................................... Normal or ventilated
Ratio (inspiration:expiration)

Normal ................................................................ 1:1, 1:2, 1:3, 1:4, 1:5
Ventilated............................................................ 1:1

Impedance Variations (

Δ

Ω

) ................................. 0.00 to 1.00

Ω

in 0.05

Ω

steps and 1.00 to 5.00

Ω

in 0.25

Ω

steps

Delta Accuracy ......................................................

±

(5 % of setting + 0.1

Ω

)

Baseline ................................................................. 500, 1000 (default), 1500, 2000

Ω

, Leads I, II, III

Baseline Accuracy ................................................

±

5 %

Respiration Lead................................................... LA or LL (default)
Apnea Selection .................................................... 12, 22, or 32 seconds (one-time events), or continuous (Apnea ON =

respiration OFF)

Power-On Default.................................................. 20 BrPM, delta 1.0

Ω

Temperature

Temperature .......................................................... 30.0

°

C to 42.0

°

C in 0.5

°

C steps

Accuracy................................................................

±

0.4

°

C

Compatibility ......................................................... Yellow Springs, Inc. (YSI) Series 400 and 700
Output .................................................................... Circular DIN 4-pin

Cardiac Output

Catheter Type ........................................................ Baxter Edwards, 93a-131-7f
Calibration Coefficient.......................................... 0.542 (0

°

C injectate), 0.595 (24

°

C injectate)

Blood Temperature ............................................... 36

°

C (98.6

°

F) to 38

°

C (100.4

°

F)

±

0.2

°

C in 1

°

C steps

Injectate Volume ................................................... 10 cc
Injectate Temperature........................................... 0

°

C or 24

°

C

Cardiac Output ...................................................... 2.5, 5, 10 liters per minute

±

7.5 %

Faulty-Injectate Curve .......................................... Waveform for simulation available
Left-to-Right-Shunt Curve.................................... Waveform for simulation available
Calibrated Pulse.................................................... 1.5

°

for 1 second

Connector .............................................................. Circular DIN 7 pin
Power-On Default ................................................. 5 liters per minute, 0

°

C injectate, 37

°

C blood temperature

Non-Invasive Blood Pressure

Pressure Units....................................................... mmHg or kPa
Manometer (Pressure Meter)

Range ................................................................. 10 to 400 mmHg
Resolution........................................................... 0.1 mmHg
Accuracy

ProSim 8 .........................................................

±

(0.5 % reading +0.5 mmHg)

ProSim 6 .........................................................

±

(1 % reading +1 mmHg)

Pressure Source

Target Pressure Range ...................................... 20 to 400 mmHg
Resolution........................................................... 1 mmHg

NIBP Simulations

Pulse................................................................... 2 mmHg max into 500 ml NIBP system
Volume of air moved........................................... 1.25 ml max
Simulations ......................................................... Systolic/diastolic (MAP)

Adult................................................................ 60/30 (40), 80/50 (60), 100/65 (77); 120/80 (93); 150/100 (117); and

200/150 (167) and 255/195 (215)

Neonatal ......................................................... 35/15 (22); 60/30 (40); 80/50 (60);100/65 (77);120/80 (93) and 150/100

(117)

Pressure variability ......................................... Systolic and diastolic pressures are variable by 1 mmHg

Repeatability ....................................................... Within

±

2 mmHg (at maximum pulse size independent of device under

test)