Automation Command and Query Reference Manual - Control Reference

Clear4

Action

Clear all calibration coefficients for Channel 4.

Description

CommonMode

Double

From -1.1 to 1.1 step 0.001

Range

The TF-DSQ fixture will calibrate probes differentially or in single-ended mode depending on the type
of probe. Differential probes allow the common mode voltage component to be applied during the DC
calibration for improved calibration accuracy in situations where probe gain or offset correction
depends on common mode components.

Description

DcCal1

Action

Initiate a DC calibration (Gain/Offset) of Channel 1.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)

Description

DcCal2

Action

Initiate a DC calibration (Gain/Offset) of Channel 2.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)

Description

DcCal3

Action

Initiate a DC calibration (Gain/Offset) of Channel 3.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)

Description

DcCal4

Action

Initiate a DC calibration (Gain/Offset) of Channel 4.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)

Description

Deskew1

Action

Initiate a deskew calibration of Channel 1.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)

Description

Deskew2

Action

Initiate a deskew calibration of Channel 2.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)

Description

Deskew3

Action

Initiate a deskew calibration of Channel 3.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)

Description

1-194

918501 RevA