JITTERMEASURE

app.SDA2.JitterMeasure

ConstantRj

Double

From 0 to 1e-006 step 1e-015

Range

ShowJitterMeasure

Bool

This control corresponds to the checkbox titled "Enable Jitter Meas" on the left of SDA2's Jitter
Measure dialog. Setting this control to True causes the "SDA Jitter" parameter table to be displayed.
That table contains at least Tj, Rj and Dj. Each parameter name is followed by more detail about how
it was computed, for example "Tj(1e-12)" means Tj was computed for the probability of error of 1e-
12; "Rj(sp)" and "Dj(sp)" mean that the jitter decomposition was done using the spectral method.
Other measurements can be added to the "SDA Jitter" parameter table (see
app.SDA2.JitterMeasure.Parameters etc).

Description

UseConstantRj

Bool

ConstantRj

Double

ShowJitterMeasure

Bool

UseConstantRj

Bool

FILTER

app.SDA2.JitterMeasure.Filter

This feature allows jitter values from TIE to be passed through a bandpass (or low pass) filter before being further
analyzed. See FilterMode, below, for more detail.

FilterMode

Enum

When FilterMode is set to On, jitter values from TIE are passed through a bandpass (or low pass)
filter before being further analyzed. If the Low Pass(band) Cutoff minus the transition width is too low,
for example if it is below 0, then a bandpass filter cannot be made; a low pass filter is automatically
used instead; we count on the use of PLL to eliminate low frequency jitter.

Description

Values

Off

On

FilterMode

Enum

HighFreqPass

Double

LowFreqPass

Double

TransitionWidth

Double