Automation Command and Query Reference Manual - Control Reference

HorPos

Double

From -0.5 to 0.5 step (8 digits)

Range

Horizontal Position of the trace, normalized to a value between -0.5 and 0.5. A value of zero is the
default, and indicates no position change relative to the source trace.

Description

HorZoom

Double

From 0.1 to 1e+006 step (8 digits)

Range

Horizontal Zoom setting. Locked to a 1, 2, 5 sequence unless VariableHorZoom is set to True .

Description

ResetZoom

Action

Resets the zoom settings to their default values.

Description

VariableHorZoom

Bool

Enable/Disable the variable Horizontal Zoom control. If enabled, the HorZoom control may be set to a
value other than the standard 1, 2, 5 sequence.

Description

VariableVerZoom

Bool

Enable/Disable the variable Vertical Zoom control. If enabled, the VerZoom control may be set to a
value other than the standard 1, 2, 5 sequence.

Description

VerPos

Double

From -1.5 to 1.5 step (8 digits)

Range

Vertical Position of the trace, normalized to a value between -1.5 and 1.5. A value of zero is the
default, and indicates no position change relative to the source trace.

Description

VerZoom

Double

From 0.1 to 100 step (8 digits)

Range

Vertical Zoom setting. Locked to a 1, 2, 5 sequence unless VariableVerZoom is set to True .

Description

HorPos

Double

HorZoom

Double

ResetZoom

Action

VariableHorZoom

Bool

VariableVerZoom

Bool

VerPos

Double

VerZoom

Double

1-467

918501 RevA