Sparse, Ssctrack, App.math.fx.operatorysetup (operator = "sparse") – Teledyne LeCroy WavePro Automation Command User Manual

Page 547: App.math.fx.operatorysetup (operator = "ssctrack")

Advertising
background image

Automation Command and Query Reference Manual - Processor Reference

VerRescale

Bool

SPARSE

app.Math.Fx.OperatorYSetup (Operator = "Sparse")

Waveform sparser, will reduce the number of points in the output waveform by skipping points in the input, and starting
at a given offset.

' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")

' Configure F1 to sparse C1 by a factor of 100
app.Math.F1.View = True
app.Math.F1.Source1 = "C1"
app.Math.F1.MathMode = "OneOperator"
app.Math.F1.Operator1 = "Sparse"
app.Math.F1.

Example

SparsingFactor

Integer

From 1 to 1000000 step 1

Range

Sets/Queries the factor by which the number of samples is reduced in the sparsing function Fx.

Description

SparsingPhase

Integer

From 0 to 0 step 1

Range

Sets/Queries the number of the first sample that will be accepted by the sparsing function Fx.

Description

SparsingFactor

Integer

SparsingPhase

Integer

SSCTRACK

app.Math.Fx.OperatorYSetup (Operator = "SSCTrack")

AutoLength

Bool

AutoLength

Bool

Bandwidth

Double

CarrierFrequency

Double

DecimateBy

Integer

FirOrIir

Enum

FMScale

DoubleLockstep

LowFreqPass

Double

TransitionWidth

Double

Type

Enum

2-48

918501 RevA

Advertising