Automation Command and Query Reference Manual - Processor Reference

eye digrams vertical information is analyzed.
It specifies the percentage of the central region of the eye (relative to 1 Unit Interval) which is used in
the analysis.

Output

Enum

Values

db

linear

pct

EYEZEROLEVEL

app.Measure.Px.Operator (ParamEngine = "EyeZeroLevel")

Aperture

Double

From 0 to 100 step 0.1

Range

For eye-diagram parameters which have an "aperture" setting, this defines the region over which the
eye digrams vertical information is analyzed.
It specifies the percentage of the central region of the eye (relative to 1 Unit Interval) which is used in
the analysis.

Description

Aperture

Double

FALLATLEVEL

app.Measure.Px.Operator (ParamEngine = "FallAtLevel")

HighPct

Double

From 10 to 95 step 1

Range

High level in percent.

Description

HighPct

Double

LevelsAre

Enum

LowPct

Double

SetLevel1090

Action

SetLevel2080

Action