Automation Command and Query Reference Manual - Control Reference

Coupling

Enum

Auxiliary trigger input connector coupling.

Description

Values

DC1M

DC50

Gnd

ProbeAttenuation

Double

From 1e-006 to 10000 step 1e-006

Range

Probe attenuation for the probe on the Auxiliary Input connector.

Description

ProbeName

String

Any number of characters

Range

Name of the probe on the Auxiliary Input connector.

Description

AUXOUTPUT

app.Acquisition.AuxOutput

Controls for the Auxilliary output BNC, which can be programmed as a simple square-wave signal source, or as a
pulse which is asserted when various events occur, including Trigger Enabled, Trigger Out, and Pass/Fail.

' Microsoft Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")

' Setup the Auxiliary output to be a squarewave with an amplitude
' of 500mV a frequency of 5kHz
app.Acquisition.AuxOutput.Mode = "Square"
app.Acquisition.AuxOutput.Amplitu

Example

Amplitude

Double

From 0.005 to 1 step 0.001

Range

Sets/Queries the amplitude of the signal on the AUX OUT connector. Note that this is the amplitude of
the signal into a 1Mohm load. Into 50 ohms the output voltage will be halved (since the source
impedance is nominally 50 ohms). Units are Volts.

Description

Amplitude

Double

AuxInCoupling

Enum

AuxMode

Enum

CalMode

Enum

Frequency

DoubleLockstep

Mode

Enum

SetToDefault

Action