Eyeheight – Teledyne LeCroy WavePro Automation Command User Manual

Page 599

Advertising
background image

Automation Command and Query Reference Manual - Processor Reference

HighPct

Double

From 60 to 90 step 1

Range

High level in percent.

Description

LowPct

Double

From 10 to 40 step 1

Range

High level in percent.

Description

EYEHEIGHT

app.Measure.Px.Operator (ParamEngine = "EyeHeight")

Aperture

Double

From 0 to 100 step 0.1

Range

For eye-diagram parameters which have an "aperture" setting, this defines the region over which the
eye digrams vertical information is analyzed.
It specifies the percentage of the central region of the eye (relative to 1 Unit Interval) which is used in
the analysis.

Description

CalcUnits

Enum

Specifies the units of the parameter readout, linear (volts), or decibels.

Description

Values

dB

linear

CursorDisplay

Enum

Set/Query the CursorDisplay cvar. This defines whether the source trace is annotated with 'Help
Markers' generated by the measurement.

Description

Values

Detailed

Off

Simple

Aperture

Double

CalcUnits

Enum

CursorDisplay

Enum

2-100

918501 RevA