Installation equipment, Installation procedure – ZOLL Thermogard XP IVTM System Operation Manual User Manual

Page 101

Advertising
background image

Accesories

-

HMIA

7

Figure 4

System Software Version

If your System does not have the compatible software version, then the correct
software will need to be installed for the HMIA to function. Failure to install the
correct software prevents the HMIA from functioning but it has no adverse affects
on the System.

Installation Equipment

The following items are necessary to install a HMIA.

1. System with correct software.

2. There are two different configurations of umbilical cables in the field. One

has a 6-pin connector and the other has an 8-pin connector. The
configuration of the HMIA corresponds to the type of connector on the
Umbilical Cable of the System being modified. HMIA part number
601033–001 is for the 8-pin circular connector and HMIA part number
601033–002 is for the 6-pin circular connector

3. TP-400 test fixture with adapter cable (Alsius Part number 500341-001)

that mates to the T1 connector of the System

4. Air trap assembly filled with water (Alsius Part number 500395-001).

5. The standard Hospital Monitor patient temperature cable that is used to

connect the Hospital Monitor to a Foley or rectal patient temperature
sensor.

6. Hospital Monitor that conforms to the YSI-400 requirements.

7. Optionally a Digital Multimeter may be used if a hospital monitor is not

available.

Installation Procedure

1. Disconnect power to the System

2. Disconnect the Umbilical Cable from the rear of the Display Housing

(See Figure 5)

600479-001 Rev 2 101/147