Language, Standby timer – ZOLL Thermogard XP IVTM System Operation Manual User Manual

Page 51

Advertising
background image

Operation

20

NOTE: Change in Design – Pump Rate

In earlier versions of the System, there was the option to set the pump rate to
either 200 or 240 ml/min. ALL Alsius catheters have been validated to
operate safely at a pump rate of 240 ml/min. This control is not longer
presented to the operator.

NOTE: Change in Design - Alarm Volume & Key Press
Volume

In earlier versions of the System, there was the option to set the alarm volume
and the key press volume. In this version:

1. The

Alarm Volume is always set to High

2. The Key Press Volume is always set to Medium

Language

This menu displays the current setting for the language used for displayed text.
The currently selected setting is highlighted.

Figure 4-16. Language Settings.

To keep the current selection, press the knob once. The current setting will not
be changed and the settings menu will be displayed.

To change the current setting, turn the knob to highlight the desired setting.
Press the knob once to enter the selection. The setting will be changed and the
settings menu will be displayed.

STANDBY Timer

The STANDBY Timer provides the User with an alarm as a reminder when the
System has been left in STANDBY for 15 or 60 minutes. The STANDBY Timer

600479-001 Rev 2 51/147

Advertising