Alarms & Corrective Actions

4

Alarm Message

Cause

Corrective Action

DOWNLOAD ERROR.
PLEASE CHECK
EXTERNAL
COMPUTER

Patient trend data download
has failed or there is no
computer connected to the
System.

Check the connection
between the computer and the
System and select “Try again.”
If problem persists, select
“Cancel” and contact ZOLL
technical support.

PATIENT DATA LOG
FULL. PLEASE
DOWNLOAD NOW

Patient trend data log is full.
Only two hours of space
remain.

Download patient trend data.
If data is not downloaded, it
will be overwritten.

Hi PATIENT
TEMPERATURE
ALARM

The patient’s temperature is
above the programmed Hi
patient temperature alarm
value.

The System alarm will reset
when the patient’s
temperature falls below the
programmed Hi patient
temperature alarm value.

Lo PATIENT
TEMPERATURE
ALARM

The patient’s temperature is
below the programmed Lo
patient temperature alarm
value.

The System alarm will reset
when the patient’s
temperature rises above the
programmed Lo patient
temperature alarm value.

PRIMARY
TEMPERATURE
PROBE
DISCONNECTED*

The primary patient
temperature probe is
disconnected.

Verify position of primary
temperature probe. Assure
that primary probe is plugged
into socket T1, or if an
interconnect cable is used, the
probe is plugged into the cable
and the cable is plugged into
T1.

PRIMARY
TEMPERATURE
PROBE DISLODGED*

Temperature output from the
primary patient temperature
probe changed by –0.2° C or
more within 10 seconds.

Replace probe in patient.

PUMP DOOR OPEN*

The pump cover is open.

Close the cover.

PUMP FAILURE*

The pump has failed.

Inspect the pump. Clear any
obvious faults and restart. If
the problem persists,
discontinue use and contact
your ZOLL service
representative (see Chapter
9
).

600479-001 Rev 2 123/147