New patient – no power down, Disposal of used components, Change the start-up kit – ZOLL Thermogard XP IVTM System Operation Manual User Manual

Page 85: Delete previous patient data

Advertising
background image

Operation

54

11.

Disconnect the serial interface cable from the System and the laptop
computer. Patient trend data downloading is now complete.

New Patient – No Power Down

Change the Start-up Kit

The system does not have to be powered down to start a new case. A new
catheter, patient temperature probe and Start-up Kit are required for each patient
– see Disposal of Used Components. To immediately start a new case without
power down.

1.

Place the System is Standby.

2.

Delete the previous Patient Data – see below.

3.

Verify the System Settings.

4.

You may select to Pre-Cool or Pre-Warm the coolant using the settings
menus.

5.

Connect the New Patient.

6.

Place the System in RUN when desired.

Delete Previous Patient Data

To delete a previous patient’s data and prepare the System for immediate use
with a new patient, without powering off the system, follow these steps in the
indicated order:

1.

Press the “Press for Enter/Menu” knob once. The menu is displayed.

2.

Turn the knob until the “New Patient” selection is highlighted. Press the
knob once to enter the selection.

3.

The New Patient menu will appear. The screen displays the message
“Select Yes to Delete Previous Patient Data.” Press the knob once to
enter “Yes.”

4.

The screen displays the message “Are you sure you want to delete the
previous patient data?” Press the knob once to enter “Yes.”

5.

The screen briefly displays the message “Previous patient data deleted”
and then displays the standby screen.

Disposal of Used Components

WARNING! SINGLE USE DEVICE–DO NOT REUSE.

Alsius catheters and System Start-up Kit components are single-use devices
and may not be reprocessed or reused. The cyclical stresses of the peristaltic
pump on the catheter and Start-up Kit cause fatigue failures.

DO NOT Use Catheters or Start-up Kits Beyond the Labeled Usage Time.
Failure of the product will result.

600479-001 Rev 2 85/147