Installing the start-up kit – ZOLL Thermogard XP IVTM System Operation Manual User Manual

Page 63

Advertising
background image

Operation

32

Setup Sensor Checks

Air Trap

There is no Start-up Kit installed or there is a large amount of air in the
Start-up Kit chamber.

Roller Pump Lid

The clear plastic roller pump lid is not closed properly.

Check Prime Switch

The Prime Switch is being depressed (as when you use it to prime the
system) or has been jammed in the ‘on” position.

Coolant

The coolant level is low. Top up the coldwell with coolant.

Installing the Start-up Kit

Hang the Saline Bag

To install the System Start-up Kit, follow these steps in the indicated order.

1.

Obtain an IV bag or bottle of sterile normal saline solution. The bag or
bottle must contain no more than 0.5 liter (500 ml) of solution. Hang the
saline container on the hook mounted on the rear of the display post.
The container hangs inside the circumference of the handle.

Figure 4-33. 500ml Saline Bag on Hook.

The Top Cover

2.

Open the top cover of the System. Open the transparent top cover of the
roller pump.

600479-001 Rev 2 63/147