Operation

40

seconds to come up to operating speed. Observe the movement of
saline until it fills the air trap and the entire length of the tubing. The
PRIME switch will only function after the system self-tests are
completed.

Figure 4-47. Use the Prime Switch to Prime the Tubing Circuit.

18.

Continue to hold the PRIME switch. When the air trap is completely filled
with saline, tap it to dislodge any remaining air bubbles. Observe the 500
ml saline bag. When bubbles are no longer seen in the saline bag,
priming is complete. Release the PRIME switch.

NOTE: Air Trap

The air trap must be completely filled for proper system operation.

19.

Turn the air trap right side up and insert it in the holder.

20.

Place the tubing to the catheter in the two notches at the front of the
console. Place the priming line and the saline return line in the channels
leading to the rear of the console. Close the top cover of the System.

CAUTION! Damage to Top Cover

Do not sit on the top cover or place heavy objects on the top cover.