Alarms & Corrective Actions

5

Alarm Message

Cause

Corrective Action

SECONDARY
TEMPERATURE
PROBE
DISCONNECTED*

The secondary temperature
probe is disconnected.

Verify position of secondary
temperature probe. Assure
that primary probe is plugged
into socket T2, or if an
interconnect cable is used, the
probe is plugged into the cable
and the cable is plugged into
T2.

SECONDARY
TEMPERATURE
PROBE DISLODGED*

Temperature output from the
secondary patient temperature
probe changed by –0.2° C or
more within 10 seconds.

Verify position of secondary
temperature probe.

TEMPERATURE
PROBES DO NOT
AGREE*

Temperature outputs from the
primary and secondary patient
temperature probes differ by
more than 2° C.

Determine the cause of the
discrepancy and correct it.
One or both temperature
probes may need to be
replaced.

SYSTEM
MALFUNCTION *

May be caused by a variety of
failures.

Discontinue use and contact
your ZOLL service
representative (see Chapter
9
).

600479-001 Rev 2 124/147