Operation

21

presents the User with the opportunity to select from the following values: No
Alarm, 15 minutes, 60 minutes.

Figure 4-17. STANDBY Timer Menu

If the system has been left in STANDBY for more than the specified time, an
alarm will sound to remind the User that the System remains in STANDBY.
Pressing the knob will reset the timer. The Standby Timer function will continue
until either:

 The System is placed into RUN.
 The STANDBY Timer Menu is used to deactivate the STANDBY

Timer.

The STANDBY Timer alert may be silenced for two (2) minutes by pressing the
mute button.

Figure 4-18. STANDBY Timer Alert