Introduction

8

WARNING! FINGER INJURIES

Be careful when inserting the pump tubing that you do not catch your fingers
with the roller.

When the System is operating DO NOT attempt to circumvent the safety
interlocks on the peristaltic pump lid. DO NOT place fingers or foreign objects
into the pump raceway when the pump is turning. The peristaltic pump has
sufficient torque to severely damage a finger.

If a tubing leak or failure occurs in the pump raceway, the saline solution will
cause corrosion in the moving parts of the rotor.

1.

Remove the pump rotor (refer to Pump Rotor removal instructions in
System Service Manual).

2.

Rinse the rotor in clean water and thoroughly dry the rotor.

3.

Apply a few drops of light machine oil to the moving parts of the rotor.

4.

Reinstall the rotor (refer to pump rotor installation instructions in System
Service Manual
).

Figure 2-5. Pump.

Prime Switch

The prime switch is located next to the pump under the top cover. The switch is
used to operate the pump to prime the tubing with sterile saline solution from the
saline source. When the switch is held down, the pump runs; when the switch is
released, the pump stops.

600479-001 Rev 2 20/147