Safety

Information

6

Shipping and Storage Conditions

When shipping or storing the System, follow these recommendations:

 Temperature range: -20ºC (-4ºF) to 60ºC (140ºF).
 Atmospheric pressure range: 50 kPa to 106 kPa.
 Do not allow liquids to spill on the System or its packaging.
 Humidity range: 10% to 90% noncondensing.
 Fragile contents, handle with care.
 Always handle and store with the top facing up.

Ignition of Flammable Anesthetic Mixtures

The System is not category AP or APG equipment and must not be used in
environments where flammable anesthetic gas mixtures are present.

Electrical Hazards

This equipment has been tested and found to comply with the EMC limits of the
international standard EN 60601-1-2. These limits are designed to provide
reasonable protection against interference in a typical medical installation. The
equipment can radiate radio frequency energy if not installed in accordance with
the instructions, and may cause harmful interference to other devices in the
vicinity. There is no guarantee that interference will not occur in a particular
installation. Always comply with the following:

 To avoid the risk of electrical shock, do not remove any panels of the

product.

 Refer servicing to qualified personnel.
 Never operate equipment with damaged power line cords.
 Refer service and repairs to a qualified technician.

CAUTION! ELECTRIC SHOCK HAZARD.

Electric shock hazard. Always turn off the System and disconnect the
power line cord from the source before performing any service or
maintenance procedures, or before moving the System.

600479-001 Rev 2 9/147