Maintenance

8

2.

Remove the cap on the coolant well and set it aside on a clean surface.

3.

Remove the front panel of the console by pulling the panel straight away
from the front of the machine (no tools are required to remove the panel).
Near the bottom left corner of the console is a drain hose held by a
clamp and the drain coupler.

Figure 8-2. Drain Hose and Drain Coupler.

4.

Place the open end of the drain hose into a suitable container or position
it above a floor sink or drain.

5.

Push the connector on the drain hose into the drain coupler on the
coolant well (refer to the following figure).

Figure 8-3. Push Hose Into Drain Coupler.

6.

As soon the connector is firmly seated in the coupler, coolant will
automatically begin to flow through the line.