Operation

45

Setup - Variations

System Setup Sequence

The System Setup sequence varies depending upon the following factors:

1.

The time from the last power down.

2.

Whether or not you used the “End Procedure” menu option when you
terminated the previous case.

Time From Last Power Down

If the System detects that it is less than 3 minutes since the last time it was
powered down it will dispense with some of its self tests depending upon the
state of the cooling engine at the time of power down. The idea being that, with
such a short time, it is more likely than not that the power down was either
unintentional (you tripped over the power cord) or an intentioned brief adjustment
that required the system to be powered off (you moved the System to a new
position).

The System will give you the option to use the last programmed settings. If you
elect to do this you may still use the menu system to change system settings in
STANDBY.

Figure 4-54. Operate at Current Settings

At power on of the system, if it has been less than 3 minutes since the last power
down, then you will be presented with the screen above.

1.

If you select “Yes” and press the knob, you will proceed directly to
STANDBY once the system self checks have completed.

2.

If you select “No” and press the knob, you will continue to be presented
with screens that allow you to select operating parameters for the
System prior to entering Standby starting at Figure 4-57 – see below.

600479-001 Rev 2 76/147