Alarms & Corrective Actions

3

Alarms and Corrective Actions

Overview

This chapter lists alarm messages that may appear on the System display during
operation. If an alarm occurs, refer to the following table for information about the
cause and corrective actions you can take to remedy the problem.

Please note that in many cases the System will automatically restart when
the condition that generated the alarm has been cleared. In other cases,
you must take corrective action and then restart the System.

Alarms and Corrective Actions

Alarm messages are listed in alphabetical order.

* Stops operation of the System.

Table 6-1. Alarm Messages, Causes, and Corrective Actions.

Alarm Message

Cause

Corrective Action

AIR TRAP FAULT*

Air has been detected in the
air trap.

Verify that the air trap is
completely filled with saline.
Inspect tubing circuit for
leaking saline. Refer to “Air
entry into the tubing circuit” in
the Preface.
If the problem persists,
discontinue use and contact
your ZOLL service
representative (see Chapter
9
).

The air trap is full of saline
but the saline level detector
indicates a fault.

Completely dry the outside of
the air trap with a towel to
remove any condensation.
Verify that the air trap is
seated firmly at the bottom of
the air trap holder.
If the problem persists,
discontinue use and contact
your ZOLL service
representative (see Chapter
9
).

COOLANT EMPTY*

The coolant well is empty.

Fill with mixture of propylene
glycol and distilled water.

COOLANT LOW

The coolant level is low.

Add distilled water until liquid
level reaches top fill line in
coolant well.