Operation

59

Cursor

The cursor is a fixed vertical line that runs through the center of both graphs.
When the temperature trend graph is displayed, you may turn the Press for
Enter/Menu knob to scroll the display to the left or right. As data scrolls under
the cursor, the top of the screen shows the time and date of the data under the
cursor.

Turn the knob clockwise to scroll to the right. As you scroll to the right, the time
and date display will indicate later data. By scrolling right to the end, you can
display the most current data.

Turn the knob counterclockwise to scroll to the left. As you scroll to the left, the
time and date display will indicate earlier data. By scrolling left to the end, you
can display the data collected when treatment began.

Status Bar

Across the bottom of the display is a status bar which displays the patient’s
temperature, the status of the System, and the target temperature for the data
point under the cursor.

The patient’s temperature and target temperature are displayed using the current
temperature notation setting (Celsius or Fahrenheit).

The System status field uses colors to display one of nine status messages that
are described in the following table.

Table 4-5. System Status Bar Messages.

Status

Message

Message

Color

Explanation

STBY

Black

The System was in standby mode (pump off).

MAX Red

MED

Red

The System was warming.

LOW Red

0

Black

The System was neither warming nor cooling.

LOW Blue

MED

Blue

The System was cooling.

MAX Blue

OFF

Black

The System was turned off.

Setting the Time Scale

The time scale displayed by the temperature trend graph can be set to any of
four intervals: 4 hours, 12 hours, 24 hours, or 72 hours. To set the time scale,
follow these steps:

1. Display the temperature trend graph.

2. Press

the

Press for Enter/Menu knob once. The screen displays a pop-

up menu with the choice “Set Time Scale” highlighted.

600479-001 Rev 2 90/147