Errors, Green led off, Flashing red led – ZOLL Thermogard XP IVTM System Operation Manual User Manual

Page 106: No temperature reported on hospital monitor, Incorrect temperature reported on hospital monitor

Advertising
background image

Accesories

-

HMIA

12

Errors

Green LED OFF

The Green LED indicates that power is present on the HMIA circuit board.

If the Green LED is not on then the LED has either burned out or the internal
power connector is dislodged. Return the HMIA to ZOLL for service.

Flashing Red LED

The Red LED flashes only when there is a problem with the CPU or the internal
self-calibration fails. Using a small solid wire (paperclip), press the reset button
(

RST

). If the Red LED continues to flash return the HMIA to ZOLL for service.

No Temperature Reported On Hospital Monitor

1. Observe Red LED.

2. If it is flashing then follow instructions for

Flashing Red LED

above.

3. If the Red LED is not flashing then disconnect the hospital monitor from

the HMIA interface cable.

4. Using a DMM measure the resistance between the two pins.

5. Using the chart

YSI-400 Temperature vs. Resistance

determine if the

resistance value measured corresponds to the temperature displayed on
the System display.

6. If the resistance value is correct then the HMIA is functioning to

specifications.

7. If the resistance value is wrong, then test Alsius interface cable for shorts

or opens.

8. If it passes the test, then press the reset button (

RST

).

9. Repeat step 5 above.

10. If the resistance value is still wrong, then disconnect the RS232 cable

from the rear of the display housing, wait for 15 seconds and then
reconnect.

11. Wait for an additional 20 seconds for the HMIA to re-establish

communication with the System

12. If the resistance value is wrong then return the HMIA to ZOLL for service.

Incorrect Temperature Reported On Hospital Monitor

1. Disconnect the hospital monitor from the HMIA interface cable.

2. Using a DMM measure the resistance between the two pins.

3. Using the chart

YSI-400 Temperature vs. Resistance

determine if the

resistance value measured corresponds to the temperature dispalyed on
the System display.

4. If the resistance value is correct then the HMIA is functioning to

specifications.

5. If the resistance value is wrong, then test Alsius interface cable for shorts

or opens.

600479-001 Rev 2 106/147