Figure 6
RS-232 Connector – Rear Display Head
9. Connect the HMIA RS232 connector, labelled
RS232
, to the mating
connector on the Rear of the Display Housing. (See Figure 6)
Figure 7
Installed HMIA – Rear Display Head
10. Select the appropriate custom Alsius interface cable that mates with the
patient cable of the hospital temperature monitor.
11. Insert the phone jack of the selected cable into the HMIA connector
labelled
T1 out
.
12. Connect the other end of the Alsius interface cable to the patient cable of
the hospital monitor.
13. Prepare the hospital monitor such that it will monitor patient temperature.