Removal of the hmia, Connection cable part numbers – ZOLL Thermogard XP IVTM System Operation Manual User Manual

Page 104

Advertising
background image

Accesories

-

HMIA

10

14. Power ON the System.

15. Verify that the Green LED of the HMIA is ON.

16. Connect a TP-400 temperature simulator to the T1 connection on the

front of the System. Set to 35

O

C.

17. Using an air trap test fixture to complete the set-up procedure,place the

System in STAND-BY.

18. Verify that the patient temperature displayed by the System is also

displayed on the hospital monitor +/- 0.1

O

C.

Note:

There may be a delay, as long as 2 minutes, before the temperature displayed
by the hospital monitor agrees, within +/- 0.1

O

C, with the temperature

displayed by the System.

19. Set the TP-400 to 38

O

C.

20. Verify that the patient temperature displayed on the System screen is

also displayed on the hospital monitor +/- 0.1

O

C.

21. Remove the TP-400 and verify that the hospital monitor recognizes the

loss of patient temperature. Typically the hospital monitor will indicate
that the probe is DISCONNECTED.

22. Reconnect the TP-400

23. Verify that the patient temperature displayed on the System screen is

also displayed on the hospital monitor +/- 0.1

O

C.

24. Power OFF the System.

25. Verify that the hospital monitor recognizes the lack of input. Typically the

hospital monitor will indicate that the probe is DISLODGED.

26. The System with the HMIA module is now available for clinical use.

Removal of the HMIA

To remove the HMIA, simply apply the installation instructions in reverse
beginning with step 12.

Connection Cable Part Numbers

The following connection cables are available to allow the connection of the
HMIA to your hospital monitor. If you have another type of connector or do not
know which connector to use, please contact ZOLL Customer Service.

Table 2 Connection Cable Part Numbers

Part

Number

Connects to

500628-001

Monitors using a 3.5mm phone jack (Bard

®

)

500629-001

Monitors using a 2.5mm phone jack (Terumo

®

)

500630-001

Monitors using a 2-pin Molex type connector with socket contacts (Mallinckrodt

®

or

Smiths Medical

®

)

500631-001

Monitors using a 2-pin Molex type connector with pin contacts (Rüsch

®

)

600479-001 Rev 2 104/147