Info, Cycle, Erase memory – ADS Environmental Services Portable FlowShark Pulse QR 775006 A2 User Manual

Page 110

Advertising
background image

Parameter Settings

6-11

Day total values menu

INFO

This menu contains the total flow values from the past 90 days, provided the
transmitter has operated without interruption for the past 90 days. If an interruption
occurs, it displays the flow totals for the uninterrupted days of operation. Reset the
total to 0 by pressing the ALT key. This reset will not impact the totalizer.

Cycle

The Portable FlowShark Pulse typically calculates flow totals at midnight (00:00 h).
However, you can modify the time at which this will occur, when necessary, through
the RUN > Day Values > Interval. However, keep in mind that modifying the time
will impact the day totals already stored in the unit’s memory.

Erase memory

This option erases the internal daily totals in memory. However, the totalizer
readings on the main screen will remain unaffected and continue to display.

1

Range of day values

2

Day values

Total day values

Time of day totalizing

Advertising