38 ads portable flowshark pulse manual – ADS Environmental Services Portable FlowShark Pulse QR 775006 A2 User Manual

Page 75

Advertising
background image

3-38 ADS Portable FlowShark Pulse Manual

When using the air-ultrasonic depth and water-ultrasonic sensors
simultaneously, use the support plate (ADS part number 5000-PPORT-PIPE-
INST-RMS2). In this configuration, the water-ultrasonic sensor mounts to the
base plate with the slotted holes at the front of the sensor. The support plate
ensures proper cable layout and placement of the water-ultrasonic sensor
behind (downstream from) the air-ultrasonic sensor.

Use the double-mounting plate to secure the air-ultrasonic sensor to the
extension plates. Install the sensor completely parallel to the flow surface.

1

Scissors jack

2

Water-Ultrasonic Velocity/Combination Sensor or Mini Water-Ultrasonic
Velocity Sensor

3

Air-Ultrasonic Depth Sensor or Mini Air-Ultrasonic Depth Sensor

Fastening the sensors to the pipe mounting system

Pipe mounting system with support plate for combined installation of a water-ultrasonic sensor and air-ultrasonic

sensor