3-30 ADS Portable FlowShark Pulse Manual

When using the water-ultrasonic velocity/combination sensor with
the integrated pressure cell, do not operate the transmitter without
the filter installed.

Removing the filter connector from the sensor connector will
automatically lock the filter connector. While a locked connector
prevents water from entering the sensor, it also disrupts air
pressure compensation. This inhibits the pressure cell from
accurately measure depth.

Make sure the air compensation hose is not hanging in the flow,
blocked, or exhibiting sharp bends. The filter must receive
continuous, unhindered air flow.

2-Wire Sensors

The Portable FlowShark Pulse supports external 4 – 20 mA 2-wire sensors for
measuring depth. The supply voltage for the sensor is 16 volts. Contact an ADS
representative for assistance in selecting an appropriate external sensor. Pre-
configured cables are available in 33-, 66-, and 98-foot (10-, 20- and 30-m) lengths.

Connect the external depth sensor to port 3 on the unit.

Sensor Cable Connection

Wire color

Function

Pin assignment

on connector

Portable FlowShark Pulse -> 2-wire 4 –
20 mA sensor

brown

white

16-volt (+)

4 – 20 mA

3

4

Pre-configured cables for 2-wire sensors

Peripherals

The Portable FlowShark Pulse has various analog and digital inputs and outputs that
allow for the connection of a variety of sensors or actuators. Refer to Appendix A,
Specifications, for an overview of these configurations.

Connect individual inputs/outputs directly to the multifunctional port using pre-
configured cables. The following table indicates the cable types that are available.
Contact an ADS representative for assistance in ordering the appropriate cable for the
particular application.