Parameter Settings

6-63

Measuring countdown Assistant

Once this first measurement is complete, create an obstruction of at least 2.6 inches
(65mm) (ADS recommends 4.7 inches (120 mm)) behind the sensor using a sandbag
or similar object.

The unit will not initiate the second depth measurement at the obstruction until h-
actual
exhibits some stability.

Creating obstruction and initiating measurements

Once the readings are stable, the Portable FlowShark Pulse will execute a new 8-
second depth measurement.

Measuring countdown for the second measurement

The unit will display the following readings after the second measurement has been
completed.

h_actual actual depth

h depth before creating an obstruction

v flow velocity reading

Q investigated flow