Calibration menu (cal) – ADS Environmental Services Portable FlowShark Pulse QR 775006 A2 User Manual

Page 195

Advertising
background image

B-8

ADS Portable FlowShark Pulse Manual

Calibration Menu (CAL)

CAL

level

calibration

option to calibrate level sensors
by entering a reference value

velocity

min. + max. value

min. value

-0.500m/s

minimum velocity possible

max. value

4.000m/s

maximum velocity possible

h_krit

H-crit:0.000

input of H-critical in [m] and V-critical in [m/s]

auto. discharge curve

yes

toggle between automatic YES/NO

V-crit determination

Manning-Strickler

kst
Ie [%]

80

Manning - Strickler coefficient

0.3

slope at measurement point

Manuell

manual entry of h and v

Assistant

calibration assistant

analog outputs

simulation

K1 V

0

direct reading output

digital outputs

turn relay On or OFF by using
arrow keys up or down

simulation

simulation of fill level with arrow keys up or down
v simulation with arrow keys right or left
output of calculated simulation value