Measurement and display functions, Display options in memory mode – ADS Environmental Services Portable FlowShark Pulse QR 775006 A2 User Manual

Page 98

Advertising
background image

Initial Startup

5-7

Measurement and Display Functions

After the system parameters are set, the Portable FlowShark Pulse will perform a
complete system reset and then begin taking measurements based on the designated
interval. The time required for taking measurements within each interval is based on
the flow and hydraulic conditions.

The number of storage events per hour is based on the number of intervals designated
within a full hour. For example, if a user configures the system at 12:17 pm to take
measurements at a 5-minute interval, the unit will store data from the first interval at
12:20 pm, the second interval at 12:25 pm, the third interval at 12:30 pm…and
continue recording every 5 minutes for a total of 12 events in one hour.

Display Options in Memory Mode

The following options are available for displaying information in memory mode:

Maintenance Turning on the Portable FlowShark Pulse for maintenance
purposes (such as for displaying data, checking sensors, or replacing the
battery pack) without modifying the parameters displays the current readings
for 4 minutes. The unit will continue to save data in the background based on
the designated measurement interval, provided the unit is configured to store
data at less than a 3-minute interval.

The unit will enter standby mode and turn off the display 4 minutes after the
last keystroke performed by the user. Then, the display will flash five times
following the designated interval. To conserve energy, the display will remain
blank while the unit continues to record data in the background at the
designated interval.

Reprogramming or Parameter Modification After you confirm a
modification by entering the PIN code, the display will go blank for a moment
and then, after the unit restarts, show the current readings for 3 minutes. The
unit will save data in the background based on the designated measurement
interval, provided the unit is configured to store data at less than a 3-minute
interval.

The unit will enter standby mode and turn off the display 4 minutes after the
last keystroke by the user. The display will flash five times following the
designated interval. To conserve energy, the display will remain blank while
the unit continues to record data in the background at the designated interval.