Installation 3-27

Removing the sensor (maintenance/control)

11. Clean or inspect the sensor as necessary. You can reinsert the sensor into the

screw connection once again as soon as the cutting ring has been replaced.
Use the rear clamp elements left on the sensor body as a detent or positioning
aid.

12. Tighten the spigot nut and M5 Allen head screws again.

Locking the sensor again after reinstallation

Cable Layout

Run the sensor cable along the pipe bottom from behind the (wedge) sensor to the
pipe wall. To minimize the risk of build-up, cover the cable with a thin, stainless
steel sheet or lay it in a small canal and then seal the canal with a permanent, elastic
material.