Maintenance and Troubleshooting

7-11

Error

Possible Reason

Potential Corrective Measures

Erroneous Readings

Unsuitable Hydraulic
Conditions at
Measurement Location

Refer to possible solutions under the Erratic
Readings
section.

External Depth (Level)
Signals

Check for proper connection.

Check for damaged or crushed cables, short
circuits, improper resistance loads, or current
customers without galvanic isolation.

Check measurement range and span.

Check input signal through the I/O menu.

Sensor

Check for proper connection.

Check for crushed or damaged sensor
cables/extension cables, short circuits, surge
arresters, or improper resistance loads.

Check the level signal, echo profile, flow
velocity signal, cable parameters, and
temperature through the I/O menu.

Verify that the sensor is installed at a location
that is free from vibration. Check sensor
installation (towards flow, horizontal
orientation), and the presence of dirt or silt.

Programming

Verify proper settings for location, such as
pipe shape and dimensions (including units),
sensor type, and sensor installation height.

Memory Card has No
Data or is Missing
Data

Memory Card

Verify whether the memory card is defective.
Check card function through the I/O >
MemoryCard > Info
menu.

Make sure the memory card is a card
supplied by ADS. Replace card from another
manufacturer with an ADS-supplied card.

Verify the method used for formatting the
card. Send cards formatted on a PC to ADS.

Transmitter

Verify that the memory card is firmly seated
(securely and completely) in the unit.

Make sure the memory card has been
seated in unit for adequate amount of time.
Data may not have been saved before card
was removed (ALT key stroke).

Programming

Verify storage is enabled through the
Memory Mode > Operation Mode > Mode.