Cal: velocity – ADS Environmental Services Portable FlowShark Pulse QR 775006 A2 User Manual

Page 157

Advertising
background image

6-58 ADS Portable FlowShark Pulse Manual

Pressing the Enter ( ) key after confirming the reference value displays an overview
screen showing all the active depth sensors and a comparison between the previous
(existing) and new offsets. The Portable FlowShark Pulse will display an error
message if the difference between the offsets is too significant and, therefore, reject
the new offset. If this occurs, repeat the calibration procedure and, if necessary, verify
the sensor installation at the location.

Depth (level) calibration screen

Implementing an offset will modify the installation height of the individual sensors in
the PAR > Level menu accordingly. Therefore, confirm that you would like to save
the offset(s) before exiting the menu by selecting YES. This will ensure the unit
accepts these adjustments. To ignore the new values and exit the adjustment
procedure, select NO. Selecting BACK will restart the procedure without accepting
the modifications.

Saving the new values

CAL: Velocity

Flow velocity screen