ADS Environmental Services Portable FlowShark Pulse QR 775006 A2 User Manual

Page 74

Advertising
background image

Installation 3-37

Snap the water-ultrasonic sensor onto the base plate with the slotted holes located
toward the rear of the sensor.

Left

Place the sensor with the cutouts (slotted holes) onto the base plate.

Center

Slide back the sensor until completely in place.

Right

Make sure the sensor locks securely and flush with the base plate.

Connecting the sensor to the base plate

Rotate the clamp handle of the scissors jack clockwise until the scissors are closed.
Then, place the entire system into the pipe and adjust and secure the system against
the pipe by rotating the clamp handle counter-clockwise.

Left

Connect the scissors jack to both sides of the final plate using the fastening clip.

Right

Close the scissors completely by rotating the clamp handle before installing in the pipe.

Preparing the mounting system for installation into the pipe

Consider the following when installing the pipe mounting system temporarily into the
pipe:

Provide adequate contact pressure against the pipe wall to prevent the
mounting system from becoming loose. This is particularly critical in large
diameter pipes and in pipes with high flow depths. For added security, install
additional stainless steel screws into the pipe wall.

Mount the system completely flush against the wall to prevent the potential for
the build-up of debris. Make sure that no gap exists between the mounting
plate and sensor or pipe bottom.

Secure the sensor cable to the downstream edge of the plate using cable ties.

Lay and secure the cable along the channel wall with clamps as necessary.

Refer to the Table of Mounting Sheets for the appropriate parts to assemble
the mounting system.