6-44 ADS Portable FlowShark Pulse Manual

Data Structure on the Memory Card

Main Directory

FLASH and PARA Subfolders

Data structure on the memory card

DATA

This folder represents the location in which the file containing the day totals
(TOTAL.TXT) is stored. Save the day totals by selecting I/O > Memory card >
Day values
. .

FLASH

This folder represents the location in which the backup file is saved (To execute the
backup, select I/0 > Memory Card > Save backup).

The name of the backup file containing the depth, velocity, flow, and temperature data
written from internal memory is always Q_H_V_T.TXT. The DIAG.TXT file
contains all messages, including error messages that may have occurred during
measurement activities. These may include the beginning and end of communication,
modem restarts, and CPU restarts following a system reset or reprogramming.

The respective message is labelled with the date and time:

>: received error/message

<: reason for error/message cleared

PARA

This folder includes all the parameter files with a date stamp. The content in these
files allows you to retrace transmitter settings corresponding to the measurement
location and parameter settings that may have been modified. The unit saves the
latest modification made within a particular day.

The filename is PA TT MM JJ .TXT, where TT = day, MM = month, and JJ = year.