12 device calibration, Sensors, Tank level zero value – ARAG Bravo 400S Crop sprayer User Manual

Page 39: Advanced setup "implement

Advertising
background image

39

10.1.12 Device calibration

Fig. 128

Allows to start calibration for the devices connected to Bravo 400S.

SENSORS

• Tank level zero value

Fig. 129

Perform the zero calibration of the level sensor in the following cases.

1 Bravo 400S displays the presence of fluid in tank, even though the tank is empty;

2 You have uploaded a tank profile (par. 10.1.9).

In order to use this menu the level sensor must be enabled (

Tank level source

, ch. 9).

Perform the calibration with empty tank.

Fig. 130

Fig. 131

1 Select the item

Tank level zero value

(Fig. 129) and press

OK

.

2 The message in Fig. 130 is displayed: follow
the instructions, then start the procedure by
pressing

OK

.

3 Press

OK

to reset the pressure sensor

residual signal.

Value out of range!

Should this alarm appear it means that
faulty values were detected: check the
sensor operation.
If problem persists, check for residual fluid
in the tank.

ADVANCED SETUP

"IMPLEMENT"

Par.

7.4

Exits the function
or data change

Confirm access
or data change

Scroll
(LEFT /
RIGHT)

Delete
selected
character

Increase /
Decrease
of data

Scroll
(UP /
DOWN)

Enter
selected
character

Advertising