Index – Yokogawa DAQWORX User Manual

Page 118

Advertising
background image

3-5

IM WX103-01E

1

2

3

4

5

Index

V

ewer

When synchronizing data files that exceed the 5 million point range, the file is used

as the start file and succeeding files are joined. If 5 million points is not exceeded
when the last file is joined, files before the start file is searched in order, and files that
would not cause 5 million points to be exceeded are joined. In this case, because
the files before the start file are also joined to the previous synchronized file, two
synchronized files for this file would exist. If you open a file before the start file and
perform synchronization, the viewer displays the first file the OS finds (either of the
two existing synchronized files). In such cases, it is recommended that the files you
wish to join are selected so that the limit is not exceeded, copied to the user’s working

folder, and synchronized there.

• Divided files can be joined up to the point in which the size of the joined file reaches 2

GB.

CF Fle Jonng Functon

Data saved to the CF card on the MX100 can be joined.
• The data that can be joined are CF files that were created on the same unit from

record start to record stop.

[File 0]

Start condition:

start

Stop condition:

divide

Measurement

Interval 1

Group of files with

file number 0

This group becomes the unit of joining

[File 1]

Start condition:

divide

Stop condition:

divide

• • •

[File 2]

Start condition:

divide

Stop condition:

divide

[File N*]

Start condition:

divide

Stop condition:

stop

*Max number of files

time

[File 0]

Start condition:

start

Stop condition:

divide

Measurement

Interval 2

[File 1]

Start condition:

divide

Stop condition:

divide

• • •

[File 2]

Start condition:

divide

Stop condition:

divide

[File N*]

Start condition:

divide

Stop condition:

stop

[File 0]

Start condition:

start

Stop condition:

divide

Measurement

Interval 3

Recording start

Recording stop

[File 1]

Start condition:

divide

Stop condition:

divide

• • •

[File 2]

Start condition:

divide

Stop condition:

divide

[File N*]

Start condition:

divide

Stop condition:

stop

• Data are joined by file groups of data with multiple measurement intervals of the same

file number.

• If a number of files equalling the number of measurement intervals of the specified file

number (if three measurement intervals are specified, the number would be 3) does
not exist in the file group, files just prior to the file group of the relevant file number are
joined.

• The resulting files that are created are limited to five million points/2 GB.
• The extension of the joined files is .mxc.
The operation is carried out by choosing File > Join in the menu. Also, when opening

data files, if you choose data saved to the CF card in the MX100 main unit, the “Join

related files” dialog box appears and the files are joined.

CF files or joined data files displayed in the active window can themselves be joined, and

Join in the menu is enabled.

If the data files displayed in the active window are not CF or joined files, Join is disabled

in the menu.

Note

When joining or performing other such processes, copy the data files onto the hard disk of the

PC from the CF card first.

3.1 Loadng Data Fles