6 setting alarms, 6 setting alarms -37, 6 settng alarms – Yokogawa DAQWORX User Manual

Page 67: Index

Advertising
background image

2-37

IM WX103-01E

1

2

3

4

5

Index

Logger

2.6

Settng Alarms

Procedure

1. Click Channel.

The Channel setup screen appears.

2. Click the Meas. Ch tab or the Math Ch tab.

3. Select the alarm type from the Type list box under Alarm 1 to Alarm 4.

When not using difference input on measurement channels, select OFF, High, or

Low. When using difference input, select OFF, dHigh, or dLow.

On math channels, select OFF, High, Low, rHigh, or rLow.

4. Click the Value box, and enter the alarm value.

5. Click the Hysteress box, and enter the alarm hysteresis value.

6. If you selected a rate-of-change alarm rHigh or rLow for the alarm type on a

computation channel, select the interval used to detect the rate-of-change from the

Interval of rate-of-change alarm list box.

Hysteresis of alarm 1

Alarm value of alarm 1

Alarm 1 type

Set Alarm 2 in a similar fashion

For math channels

Rate-of-change interval when the Type is set to rHigh or rLow

Hysteresis of alarm 1

Alarm value of alarm 1

Alarm 1 type

Hysteresis of alarm 2

Alarm value of alarm 2

Alarm 2 type

Click here.

Click here to set the alarm

per channel

For measurement channels

Click here to set the alarm

per channel

Scroll the screen and set alarms 3 and 4 in the same manner