Chapter 4 calibrator, 1 connecting the mx100 data acquisition unit, Chapter 4 calbrator – Yokogawa DAQWORX User Manual

Page 145: 1 connecting the mx100 data acquisition unit -1, 1 connectng the mx100 data acquston unt, Index

Advertising
background image

4-1

IM WX103-01E

1

2

3

4

5

Index

Cal

brator

4.1

Connectng the MX100 Data Acquston Unt

Notes on Usng the Calbraton Software

• If another user or another software program is connected to the MX100, drop the

connection before connecting with the Calibration Software.

• If you change the module configuration before connecting with the Calibration

Software, carry out system configuration on the Logger (see section 2.2, “Configuring
the System”).

• When calibrating, do not perform other operations. Use extra caution in the operation,

because the module may malfunction if you execute a wrong operation while
calibration is in progress.

Procedure

1. Start the Calibrator.

The Unit Information window opens.

2. In the IP Address/Host Name box, enter the IP address or host name of the MX100

to be calibrated.

3. Click Connect.

Enter the host name or IP address of the MX100 to be calibrated.

Execute connection

When the connection is established, information about the unit’s module
configuration is displayed. If you click the Check button, the 7-segment LED on the
main module of the connected unit shows the text “—CALL—” flowing from right to
left.

To disconnect the connected MX100

To turn on the 7-segment

LED on the connected

MX100 to check which unit

is the one connected

Information about the connected MX100

Modules attached to the

connected MX100 Data

Acquisition Unit

Chapter 4

Calbrator

Advertising