3-12

IM WX103-01E

Clppng of Waveforms

By default (clip OFF), the waveform is not displayed when the measured/computed value

exceeds the minimum/maximum value of the scale (see the lower left figure). When clip

is turned ON, values that are smaller than the minimum value of the scale are displayed

as the minimum value and the values that are larger than the maximum value of the

scale are displayed as the maximum value.

Clip OFF

Clip ON

Expandng or Reducng the Tme Axs on the Waveform Dsplay

On the waveform display window, you can click the expand/reduce icon to expand or

reduce the time axis in the range of 20 to 1/1000 times for each waveform display area.

Magnification

Time axis expand/reduce icon

Addng Marks

In addition to the marks placed using other programs such as the Logger, you can add

marks at the positions where you click (displays a cursor) on the waveform display

area. You can enter a text to be attached to the mark (“Mark” by default) using up to 15

characters. You can also select Left, Center (default), Right, or Flag (small mark without

text) for the Type. For types other than Flag, the specified string and the relative or

absolute time at the mark position are displayed.

You can set the string and type in the Mark dialog box that opens when you click the

Append Mark icon (or choose Append Mark from the Edit menu). In the Mark dialog box,

you can select whether to add the mark to the waveforms of all groups (default) or only

the waveforms of the displayed group.

To delete specific displayed marks, select the range using cursors, and then choose

Delete Mark from the Edt menu. To delete all the marks added using the Viewer, choose

Reset Mark from the Edt menu.

Left end

Mark display icon

Center

Flag

Display the cursor

or

Marks added on the Viewer are green,

marks added created through

computation are yellow, and marks

created using the Logger or MX100

Standard Software are orange.

Double-click the mark to change the settings

Grd Densty

You can change the displayed grid density. The available settings are standard, dense

grid 1, or dense grid 2. For instructions, see section 2.10, “Starting/Stopping Data

Monitoring and Displaying the Monitor.”