Setting the unit, tag number, and tag comment, Setting the rjc (when the input mode is set to tc), Rrjc (tc)) -16 – Yokogawa DAQWORX User Manual

Page 46: Setting the unit, tag number, and tag comment -16, Set to something other than di) -16, Settng the unt, tag number, and tag comment, Settng the rjc (when the input mode is set to tc)

Advertising
background image

2-16

IM WX103-01E

Settng the Remote RJC Reference Channel (Vald Only When the Input Mode Is
Set to RRJC (TC))

12. Select the reference channel for the remote RJC from the Ref. Ch. list box under

RRJC.

Note

If the reference channel setting for difference input or the reference channel setting for the

remote RJC is inappropriate when switching from the Channel screen to the Display or Monitor

screen, a Channel dialog box opens (see the figure below). To disable the measurement

channels with inappropriate settings (clear the Monitor check box on the Channel screen) and

move to the Display or Monitor screen, click Yes. To change the settings of the inappropriate

measurement channels without moving to the next screen, click No.

Settng the Unt, Tag Number, and Tag Comment

13. Click the Unit, Tag No., and Tag Comment text box and enter the unit, tag number,

and tag comment, respectively.

Settng the Tme Constant of the Frst-Order Lag Flter (Vald Only When the
Input Mode Is Set to Somethng Other than DI)

14. Select the time constant from the Flter list box.

Settng the Burnout (Vald Only When the Input Mode Is Set to TC)

15. Select the direction in which the measured value is set off the range (+OVER or –

OVER) when a burnout occurs from the Burnout list box.

Settng the RJC (When the Input Mode Is Set to TC)

16. Click the Type box under RJC and select Internal or External.

17. If you select External, click the Volt box under RJC and enter the compensation

voltage.

Enter the tag comment

(up to 30 characters)

Enter the unit (up to six characters, only

when the Use check box of Scale is selected)

Enter the tag No.

(up to 15 characters)

Select the reference channel for the remote RJC

(valid only when the input mode is set to RRJC (TC))

Set the RJC

(valid only when the input

mode is set to TC)

Set the burnout

(valid only when the input mode

is set to TC)

Set the time constant of the first-order

lag filter

(when the input mode is other than DI)