Index – Yokogawa DAQWORX User Manual

Page 51

Advertising
background image

2-21

IM WX103-01E

1

2

3

4

5

Index

Logger

MX100

Thermocouple*

Copper wire

Copper wire

Thermocouple*

Thermocouple*

Relay terminal (to be furnished by the user)

Reference channel

Input terminal

* Use the same type of thermocouples.

Settng the Measurement Span

Set the minimum and maximum values of the range that is actually measured within the

measurable range.

Settng the Scale

Set this item when linearly scaling the measured values. Set the scale by entering the

maximum and minimum values corresponding to the maximum and minimum values of

the measurement span and selecting the decimal point position of the scaled value (see

the figure below). The selectable range of the scale is -30000 to 30000. The decimal

point position can be set to 4 (XXXXX), 3 (XXXXX), 2 (XXXXX), 1 (XXXXX), or 0 (XXXXX).

Also, you can specify units for the scale value. You can set the units using up to six

characters.

0

3[V]

0

100

Span minimum

Scale minimum

Span maximum

Scale maximum

Settng the Reference Channel for Dfference Input

Set this item when making the difference between the measured value of the channel

and the measured value of the reference channel the measured value (referred to as

difference computation). The reference channel can be set to a channel that is scaled.

Difference computation can be executed even if the measurement range of its own

channel and that of the reference channel are not the same.

If the decimal point position setting is different between its own channel and the reference

channel, the decimal point position of the measured value of the reference channel is

matched to that of its own channel, and the difference is computed.

Example: When the measured value of the channel set to difference input is 10.00 and

the measured value of the reference channel is 100.0

The computed result is 10.00 – 100.0 = –90.00.

Set the reference channel for difference input to a measurement channel that has the

Monitor check box selected.

Settng the Remote RJC Reference Channel (Vald Only When the Input
Mode Is Set to RRJC (TC))

This item sets the reference channel when the input mode is set to RRJC (TC).

Set the reference channel for remote RJC to a channel that meets the three conditions

below.

2.4 Settng the Measurement Condtons (Measurement Channel Settngs)