Index, Explanaton – Yokogawa DAQWORX User Manual

Page 49

Advertising
background image

2-19

IM WX103-01E

1

2

3

4

5

Index

Logger

Explanaton

Input Mode and Measurement Range

The measurement channel settings differ depending on the combination of the

MXLOGGER software version and the MX100 hardware style number. The table below

shows the relationship between MXLOGGER measurement channel settings (by

software version) and the MX100 hardware style.

MXLOGGER

MX100 Hardware

Style Number

*2

Notes

Software verson

*1

Connectable module types
or settng tems

S3

S2

S1

R2.05 or later

30ch module
MX110-VTD-L30

Yes

No

No

GOST range

Yes

No

No

Can be set only for MX110-UNV-M10,
MX110-UNV-H04, MX110-V4R-M06, and
MX110-VTD-L30

Yes: Setting and operation allowed

No: Setting possible, but not operation

*1 To check the version, see section 2.16, “Displaying the Software Version and Other

Information.”

*2 The style number is printed on the MX100 main module name plate.

On the channels of the universal input module, select from the modes below according

to the input signal. For details of the measurable range, measurement accuracy, and

resolution, see chapter 4, “Specifications” in the MX100 Data Acquisition Unit User’s

Manual (in the manual CD-ROM)

• VOLT (DC voltage)

Select the measurement range from 20 mV, 60 mV, 200 mV, 2 V (default), 6 V, 20 V,

100 V, 60 mV (HQ), 1 V (HQ), and 6 V (HQ). HQ denotes high resolution.

Note

Current input is possible by attaching a shunt resistor to the input terminal and converting to voltage

input. The shunt resistors in the table below are available for purchase separately. For example, a
250 Ω shunt resistor is used to convert the signal in the range of 1 to 5 V for 4 to 20 mA input.

Part Name

Model

Resstance

Shunt resistor

415920

250 W ± 0.1%

(for clamp terminal)

415921

100 Ω ± 0.1%

415922

10 Ω ± 0.1%

Shunt resistor

438920

250 Ω ± 0.1%

(for screw terminal)

438921

100 Ω ± 0.1%

438922

10 Ω ± 0.1%

• TC (Thermocouple)

Select the thermocouple type (referred to as Range in the setup) from Type-R (default), Type-S,

Type-B, Type-K, Type-E, Type-J, Type-T, Type-N, Type-W, Type-L, Type-U, and KpvsAu7Fe,

PLATINEL, PR40-20, NiNiMo, WRe3-25, W/WRe26, Type-N(AWG14), and Type-XK.

• RTD1 (resistance temperature detector, measurement current: 1 mA) on the 4-CH,

High-Speed Universal Input Module

Select the RTD type (referred to as Range in the setup) from Pt100 (default), JPt100,

HQ Pt100, HQJPt100, Ni100:SAMA, Ni100:DIN, Ni120, RN Pt100, RN JPt100 and

Pt100G. HQ and RN denote high resolution and noise resistance, respectively.

2.4 Settng the Measurement Condtons (Measurement Channel Settngs)

Advertising
This manual is related to the following products: