T

roubleshoot

ng

MX100 cannot be connected usng the Calbrator

Probable Cause

Correctve Acton

Reference Page

Attempting to make multiple connections.
Another software program is already
connected.

Exit all other software programs.

-

The connected nput/output module s not detected.

Probable Cause

Correctve Acton

Reference Page

Module connection or module startup
error.
Attached the module while the power was
ON.

Turn OFF the power. Detach the input/output module once
and
attach it again.

*

Carried out an incorrect calibration.

Recalibrate.

4-1

* See the

MX100 User’s Manual (IM MX100-01E).

Alarms are not output.

Probable Cause

Correctve Acton

Reference Page

There is a problem in the alarm setting.

Both the alarm and output relay must be set appropriately.
Make the alarm and output relay settings appropriate.

2-37~2-42

Current s not output from the analog output module, and no output comes from the

PWM output module.

Probable Cause

Correctve Acton

Reference Page

External power supply required.

Connect an external power supply.

*

Transmsson output s not output from the analog output module, nor from the PWM

output module.

Probable Cause

Correctve Acton

Reference Page

“Transmit not enabled” Enable it.

Enable it.

2-50

No output from the PWM output module

Probable Cause

Correctve Acton

Reference Page

The output current limit (1A) has been
exceeded. Turn OFF the external power
supply, check the load, power cycle.

Once the limit is exceeded the protection circuit activates,
and you must power cycle the external power supply or it
will not recover.

-

No output from the PWM output module

Probable Cause

Correctve Acton

Reference Page

The fuse is blown.
(4A/module limit has been exceeded)

Servicing required.

-

The CF card s not detected.

Probable Cause

Correctve Acton

Reference Page

There is a problem with the CF card.

Replace the CF card.
Eject and format the CF card, then insert it again.

*