Yokogawa DAQWORX User Manual

Page 50

Advertising
background image

2-20

IM WX103-01E

• RTD2 (resistance temperature detector, measurement current: 2 mA) on the 4-CH,

High-Speed Universal Input Module

Select the RTD type (referred to as Range in the setup) from Pt100 (default), JPt100,

HQ Pt100, HQ JPt100, Pt50, Cu10:GE, Cu10:L&N, Cu10 :WEED, Cu10:BAILEY, J263B,

Cu10 a=0.00392, Cu10 a=0.00393, Cu25, Cu53, Cu100, Pt25, HQ Cu10:GE, HQ Cu10:

L&N, HQ C10:WEED, HQ Cu10:BAILEY, RN Pt100, RN JPt100, Cu100G, Cu50G, and

Cu10G. HQ and RN denote high resolution and noise resistance, respectively.

The RTD2 selection is not available on the 10-CH, Medium-Speed Universal Input

Module or the Six-Channel, Medium-Speed, Four-Wire RTD Resistance Input Module.

• RTD1 (resistance temperature detector, measurement current: 1 mA) on the 10-CH,

Medium-Speed Universal Input Module/Six-Channel Medium-Speed Four-Wire RTD

Resistance Input Module

Select the RTD type (referred to as Range in the setup) from Pt100 (default), JPt100,

HQ Pt100, HQ JPt100, Ni100:SAMA, Ni100:DIN, Ni120, Pt50, Cu10:GE, Cu10:L&N,

Cu10 :WEED, Cu10:BAILEY, J263B, Cu10 a=0.00392, Cu10 a=0.00393, Cu25,

Cu53, Cu100, Pt25, HQ Cu10:GE, HQ Cu10:L&N, HQ Cu:WEED, HQ Cu10:BAILEY,

Pt100G, Cu100G, Cu50G, and Cu10G. HQ denotes high resolution.

• RTDEx (resistance temperature detector, measurement current: 0.25 mA) on the Six-

Channel, Medium-Speed Four-Wire RTD Resistance Input Module

Select the RTD type (referred to as Range in the setup) from Pt500 (default setting) or

Pt1000.

• OHM (resistance) on the Six-Channel, Medium-Speed Four-Wire RTD Resistance

Input Module

Choose 20 ohm (measurement current: 1 mA), 200 ohm (measurement current: 1

mA), or 2 kohm (measured current: 0.25 mA).

• DI (Digital Input)

On the channels of the 10-CH, High-Speed Digital Input Modules, only the DI (digital

input) setting is available (no input mode selection). With numerical input module

channels, select LEVEL (D10) (voltage input) for -D05, or CONTACT (D10) (contact

input) for the measurement range to match the input. -DI24: LEVEL (24 V) only.

On the 4-CH, High-Speed Universal Input Module, select LEVEL (voltage input) or

CONTACT (HS) (contact input) according to the input.

On the 10-CH, Medium-Speed Universal Input Module or the Six-Channel Medium-

Speed Four-Wire RTD Resistance Input Module, select LEVEL (voltage input) or

CONTACT (SC) (contact input) according to the input.

• STR (Strain)

On the 4-CH Medium-Speed Strain Input Modules, select a measurement range of

2000 µSTR, 20000 µSTR (default value), or 200000 µSTR.

• RRJC(TC) (Remote RJC)

In measuring temperature with thermocouples within the same unit, when the item

to be measured is located at a great distance, you can setup relay terminals near

the item, measure the temperature of the relay terminal section using thermocouples

(reference channel), and use the resultant value as the reference junction

compensation value for the temperature measurement. By connecting a copper wire

between the relay terminal and input terminal of the input module, and a thermocouple

between the DUT and relay terminal, you can measure the temperature of the DUT

without the need for a large amount of expensive thermocouples.

2.4 Settng the Measurement Condtons (Measurement Channel Settngs)