1 setting the type of input and recording span, Chapter 5, Setting the type of input and recording span -1 – Yokogawa DR240 User Manual

Page 104: 10:volt/2v, 01:volt/2v, Span=–_200.0/ ˚c, Set ok

Advertising
background image

5-1

IM DR232-01E

5

Setting the Input Type/Recording Span/Linear Scaling

5.1

Setting the Type of Input and Recording Span

D I S P

M O D E

P R I N T

F U N C

I N S

D E L

R A N G E A L A R M

S E T

C H A R T

E N T E R

F E E D

M. F U N C 1

M. F U N C 2

RECORD

ALARM

CHART

KEYLOCK

R E C O R D

001-10:VOLT/2V

SKIP VOLT TC RTD DI DELTA SCL

SPAN=–2.0000/ 2.0000V

E S C

1 Press the RANGE key to enter the RANGE menu.
2 Select/set

using the

keys.

3 To escape from a lower menu, press the “MODE (ESC)” key. Its main menu will appear,

although new settings/selections will not be kept.

4 New settings/selections will be kept when you reach the final display of that setting.

001-01:VOLT/2V

>Select Channel No.

001-01:VOLT/2V

>Select Channel No.

001-01:VOLT/2V

001-01:VOLT/2V

20mV 60mV 200mV 2V 6V 20V 50V

SPAN=–2.0000/_2.0000V

>Span limit(–2.0000~2.0000V)

***SET OK***

001-01:SKIP

***SET OK***

001-01:TC/R

001-01:TC/R

R S B K E J T N W L U KP

SPAN=_ _ _ _0.0/1760.0˚C

>Span limit(0.0~1760.0˚C)

***SET OK***

001-01:RTD/PT1

001-01:RTD/PT1

PT1 PT2 JPT1 JPT2 PT50 NI1 NI2 NI3

CU1 CU2 CU3 CU4 PT1S PT2S JPT1S

JPT2S J263B

SPAN=–_200.0/_ _600.0˚C

>Span limit(–200.0~600.0˚C)

***SET OK***

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ESC

ENTER

ESC

ESC

ESC

To the next page

(Main menu)

(Lower menu)