5 connecting the interface cables, Connecting the interface cables -14, Gp-ib – Yokogawa DR240 User Manual

Page 59

Advertising
background image

3-14

IM DR232-01E

3.5

Connecting the Interface Cables

When connecting a personal computer to the instrument via a communication interface, observe
the following:

GP-IB

The GP-IB connector of the GP-IB communication module is a 24-pin connector of IEEE St’d
488-1978. Only use cables that conform to IEEE St’d 488-1978 as a communication cable.

Connection Procedure

Connect the cable as shown in the figure below.

GP-IB module

Ferrite core

Personal
computer

Ferrite core

Function
grounding
terminal

•When connecting the cable, take note of the following.

• To reduce noise, use two ferrite cores (e.g., ZCAT 3035-1330 from TDK) at both ends of the

interface cable as shown above.

• Securely tighten the screws which fasten the GP-IB cable connector.
• If more than one equipment is connected, the same addresses cannot be assigned to different

equipment.

• Use only cables of 2 m or less to interconnect each equipment.

How to Assign an Address

The address can be assigned easily by turning the rotary dipswitch as shown in the figure below.
Any address can be set from “0” to “15”; the characters “A” to “F” on the dipswitch correspond to
the address “10” to “15” respectively.

0 12

3

4

5

678

9

F

E

D

C

B A

0 12

3

4

5

678

9

F

E

D

C

B A

Advertising