List of menus and set-up data, Measurement condition settings, Chart speed settings – Yokogawa DR240 User Manual

Page 14: Recording settings, Display settings, Alarm settings

Advertising
background image

13

IM DR232-01E

List of Menus and Set-up Data

The following is a list of set-up data, procedures to switch to different setting modes, and setting
menu.

Measurement Condition Settings

Parameters

Procedure

Selecting menu

Reference

Input type, span, linear

RANGE key

001-01:VOLT/2V

Chapter 5

scaling*

1

Units

SET key*

2

SET=UNIT

Section 6.1

Moving average

Press the SET key for about three seconds*

2

SET=MOVE AVE

Section 6.8

Measurement cycle*

1

Turn ON power while pressing the DISP key

SET UP=

Section 10.1

SCAN INTVL

A/D integration time*

1

Turn ON power while pressing the DISP key

SET UP=A/D INTG Section 10.3

Filter*

1

Turn ON power while pressing the DISP key

SET UP=FILTER

Section 10.3

*1: Make sure that the total number of setting changes, including calibrations and restructuring, does not

surpass 100000.

*2: Procedure varies according to the menu configuration of the SET key (see section 10.9).

Chart Speed Settings

Parameters

Procedure

Selecting menu

Reference

Chart speed 1

CHART key

CHART

Section 6.2

Chart speed 2

Press the SET key for about three seconds* SET=CHART2

Section 6.2

*: Procedure varies according to the menu configuration of the SET key (see section 10.9).

Recording Settings

Parameters

Procedure

Selecting menu

Reference

Logging/Analog trend

SET key*

SET=SYSTEM

Section 6.1

switch, dot-printing cycle
Recording channel

SET key*

SET=TREND

Section 6.1

Recording zone

Press the SET key for about three seconds* SET=ZONE

Section 6.3

Partial compression

Press the SET key for about three seconds* SET=PERTIAL

Section 6.3

Tag

Press the SET key for about three seconds* SET=TAG

Section 6.4

Channel to digital print, Press the SET key for about three seconds* SET=DIGITAL PR

Section 6.4

number of rows to print
Channel to manual print Press the SET key for about three seconds* SET=MANUAL PR Section 6.4
Alarm print

Press the SET key for about three seconds* SET=ALARM PR

Section 6.5

Channel to print scale

Press the SET key for about three seconds* SET=SCALE PR

Section 6.6

values
Channel to list print

Press the SET key for about three seconds* SET=LIST PR

Section 6.6

Items to list print

Press the SET key for about three seconds* SET=LIST FMT

Section 6.6

Message

Press the SET key for about three seconds* SET=MESSAGE

Section 6.7

Header

Press the SET key for about three seconds* SET=HEADER

Section 6.7

Title

Press the SET key for about three seconds* SET=TITLE

Section 6.7

Interpolation

Press the SET key for about three seconds* SET=INTERPOL

Section 6.8

Adjust dot-printing

Turn ON power while pressing the DISP key

SET UP=PRN ADJ

Section 10.1

position
Recording format

Turn ON power while pressing the DISP key

SET UP=RECORD

Section 10.2

Dot-print color

Turn ON power while pressing the DISP key

SET UP=COLOR

Section 10.6

*: Procedure varies according to the menu configuration of the SET key (see section 10.9).

Display Settings

Parameters

Procedure

Selecting menu

Reference

Switch display

DISP key and MODE key

--------

Chapter 4

Display update interval

Turn ON power while pressing the

SET UP=DISPLAY

Section 10.10

DISP key

Alarm Settings

Parameters

Procedure

Selecting menu

Reference

Alarm, alarm output

ALARM key

001-01:1/OFF

Section 8.1

relay
Alarm

Turn ON power while pressing the DISP key

SET UP=ALARM

Section 10.3

interval/hysteresis/hold
Execute alarm

FUNC key*

ALARM ACK

Section 9.3

acknowledge
Reset alarm

FUNC key*

ALARM RST

Section 9.3

Clear alarm buffer

Press the FUNC key for about three seconds* ALM BUF CLEAR

Section 9.4

*: Procedure varies according to the menu configuration of the FUNC key (see section 10.8).