Chapter 14 specifications, 1 specifications of dr232/dr242 and ds400/ds600, Specifications of dr232/dr242 and ds400/ds600 -1 – Yokogawa DR240 User Manual

Page 246

Advertising
background image

14-1

IM DR232-01E

Specifications

14

14.1 Specifications of DR232/DR242 and DS400/

DS600

Style Numbers DR232/DR242: S8

DS400/DS600: S8

Connecting Subunits to the DR232/DR242

Types of Subunits

Subunit DS400: four modules
Subunit DS600: six modules

Number of Subunits which can be connected

Up to six units can be connected (either DS400 or DS600)

Connection method

Using dedicated cables

Available Modules

Main Unit

Communication interface modules : Selectable from GP-IB and RS-232-C
DI/DO modules

: Alarm 2-ch output (transfer contact), fail output module,

chart-end output and input signal for remote control (12
points)

Alarm modules

: 4-ch output(transfer contact) or 10-ch output (make contact)

• Number of modules which can be connected: 1

Communication module + DI/DO module + alarm module: not more than one

Subunit (DS400/DS600)

Input modules

: universal (DC voltage, thermocouple, RTD, contact), DCV/TC/DI

Alarm modules

: 4-ch output (transfer contact) or 10-ch output (make contact)

DI/DO modules : Alarm 2-ch output (transfer contact), fail output module, chart-end output and

input signal for remote control

• Number of modules which can be connected to one DS400: 4

Input module + alarm module + DI/DO module: max. four

• Number of modules which can be connected to one DS600: 6

Input module + alarm module + DI/DO module: max. six

* Only one DI/DO module can be connected to one system.

Input

Measurement range

Refer to 14.2 on page 14-7.

Measurement interval

Selectable from 0.5, 1, 2, 3, 4, 5, 6, 10, 12, 15, 20, 30, 60s; max. 500ms/300ch
(with sub-unit connected, filter: OFF, A/D integration period of 20ms or 16.7ms)

A/D integration time

Selectable from 20ms (50Hz), 16.7ms (60Hz), 100ms (10Hz) selectable, or automatically
switched between 50/60Hz. (Except, auto switch does not function on the subunit (DS400/
DS600) of a DC power supply model. Selecting “AUTO” will set the A/D integration time to 20
ms (50 Hz).)

Recording

Recording method

Raster scan method, 10-color wire dot recording

Number of recording points

Measurement results:

Max.300 points

Recording paper

Ordinary recording chart, Z-fold chart (total width 342.5mm, length 30m)
Effective recording width: 250mm (for dot recording)

Recording colors

Analog trend mode

Dot recording: purple, red, green, blue, brown, black, navy blue, yellowish green, redish purple,
orange (can be specified for each channel)
Digital printout: black
Alarm printout: red (however, alarm release mark: blue

Advertising
This manual is related to the following products: